An artefact desynchroniser for use with visual evoked potential stimulators

Autor: L. B. Brown, Anthony T. Barker, A. J. Wilson
Rok vydání: 1999
Předmět:
Zdroj: Medical & Biological Engineering & Computing. 37:400-402
ISSN: 1741-0444
0140-0118
DOI: 10.1007/bf02513319
Popis: When recording visual evoked potentials elicited by a CRT raster display, driven from a frame-synchronised stimulus generator, artefacts caused by magnetic fields originating from the CRT field coils may seriously contaminate the recordings. This interference cannot be removed by signal averaging as it is time locked to the frame rate of the CRT. The circuit described here effectively desynchronises the interference, enabling the signal averaging process to reduce the artefact without the bandwidth restriction and resultant distortion of the recording which would be caused by lowpass filtering.
Databáze: OpenAIRE