An artefact desynchroniser for use with visual evoked potential stimulators
Autor: | L. B. Brown, Anthony T. Barker, A. J. Wilson |
---|---|
Rok vydání: | 1999 |
Předmět: |
Signal processing
Computer science Acoustics Bandwidth (signal processing) Biomedical Engineering Signal Processing Computer-Assisted computer.file_format Frame rate Electronics Medical Computer Science Applications Electrophysiology Electronic engineering Evoked Potentials Visual Humans Dither Signal averaging Raster graphics Evoked potential computer Electronic circuit |
Zdroj: | Medical & Biological Engineering & Computing. 37:400-402 |
ISSN: | 1741-0444 0140-0118 |
DOI: | 10.1007/bf02513319 |
Popis: | When recording visual evoked potentials elicited by a CRT raster display, driven from a frame-synchronised stimulus generator, artefacts caused by magnetic fields originating from the CRT field coils may seriously contaminate the recordings. This interference cannot be removed by signal averaging as it is time locked to the frame rate of the CRT. The circuit described here effectively desynchronises the interference, enabling the signal averaging process to reduce the artefact without the bandwidth restriction and resultant distortion of the recording which would be caused by lowpass filtering. |