Autor: |
Coxx, Don C., Furman, William B. |
Zdroj: |
Journal of Pharmaceutical Sciences; August 1984, Vol. 73 Issue: 8 p1125-1127, 3p |
Abstrakt: |
Recently marketed glass vessels that are uniform and pass USP specifications were compared with uniform plastic vessels that also pass USP specifications. Two lots of prednisone tablets, Tablet 1 and Tablet 2, were tested in both types of vessels. Tablet 1 gave higher results (+12.7% of label claim) in glass vessels at 50rpm but gave equivalent results in either vessel at 75rpm. Tablet 2 gave equivalent results in either vessel at 50rpm but gave higher results (+22% of label claim) in glass vessels at 75rpm. The type of vessel used to obtain dissolution results for tablets should be specified. |
Databáze: |
Supplemental Index |
Externí odkaz: |
|