Impact of latest generation cardiac interventional X-ray equipment on patient image quality and radiation dose for trans-catheter aortic valve implantations

Autor: Gislason-Lee, Amber J., Keeble, C., Malkin, C.J., Egleston, D., Bexon, J., Kengyelics, S.M., Blackman, D., Davies, A.G.
Jazyk: angličtina
Rok vydání: 2016
Předmět:
Druh dokumentu: Článek
DOI: 10.1259/bjr.20160269
Popis: Yes
Objectives: This study aimed to determine the impact on radiation dose and image quality of a new cardiac interventional X-ray system for trans-catheter aortic valve implantation (TAVI) patients compared to the previously-used cardiac X-ray system. Methods: Patient dose and image data were retrospectively collected from a Philips AlluraClarity (new) and Siemens Axion Artis (reference) X-ray system. Patient dose area product (DAP) and fluoroscopy duration of 41 patient cases from each X-ray system were compared using a Wilcoxon test. Ten patient aortograms from each X-ray system were scored by 32 observers on a continuous scale to assess the clinical image quality at the given phase of the TAVI procedure. Scores were dichotomised by acceptability and analysed using a Chi-squared test. Results: Significant reductions in patient dose (p