Effect of prolonged pressure equalization on final drifting during pressure wire studies

Autor: Chien-Boon Jong, Tsui-Shan Lu, Lin Lin, Tsung-Yan Chen, Min-Tsun Liao, Jui-Cheng Kuo
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Scientific Reports, Vol 14, Iss 1, Pp 1-8 (2024)
Druh dokumentu: article
ISSN: 2045-2322
DOI: 10.1038/s41598-024-62440-1
Popis: Abstract Pressure drifting is a troublesome error in invasive coronary function tests. This study aimed to evaluate the relationship between prolonged and short-time pressure equalizations in pressure drifting. Pressure drifting was defined as the pressure gradient between the mean pressure of the distal wire sensor (Pd) and aortic pressure (Pa) when the wire was withdrawn to the tip of the guiding catheter. Significant drifts 1 and 2 were defined as the absolute values of pressure gradients > 2 and > 3 mmHg, respectively. A logistic regression model was used to evaluate the associations between prolonged pressure equalization and each pressure drifting. The prolonged pressure equalization strategy was associated with a lower incidence of drift 1 than the short-time pressure equalization strategy (6.84% vs. 16.92%, p