Autor: |
K.M. El-Kourghly, W. El-Gammal, Mohamed M. Damoom, Abdulsalam M. Alhawsawi, M. Abdelati, H.I. Khedr, R. Abouzaid |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Heliyon, Vol 10, Iss 7, Pp e28498- (2024) |
Druh dokumentu: |
article |
ISSN: |
2405-8440 |
DOI: |
10.1016/j.heliyon.2024.e28498 |
Popis: |
In this work, an absolute method to calibrate an α-spectrometer is proposed taking into account the Source-to-Detector, and lateral distances due to eccentric source distribution. An analytical formula to calibrate an α-spectrometer is derived and the Simpson's integration method was utilized to solve these equations in its integral form numerically using a written C computer code. The general Monte Carlo N-particle code, MCNP as well as experimental measurements for some standard α-emitters are used to benchmark the proposed method. An agreement was found between the efficiency results calculated by MC and the proposed method with a maximum relative difference of about 0.5%. While, experimental measurement of α-emitters activity employing the proposed method differs by about 1.65% from the certified values. Accounting for the man made error allows to accurately quantify the assayed sample. Therefore, the inaccuracy in the efficiency results due to non-accurate inputs pertained to the source, and detector radii, Source–Detector distances, and eccentric source distribution are investigated in the Source-to-Detector distance range of (4 to 44 mm). The results show that a difference of ±1% in the detector radius, and Source-to-Detector distance than the normal values yields a relative difference of about ±2%, while a difference of ±50% in the source radius or source lateral distance from detector symmetry axis could only yields inaccuracy of less than ±3% in the efficiency results. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|