Effect of Novel Design Modifications on Fibrotic Encapsulation: An In Vivo Glaucoma Drainage Device Study in a Rabbit Model

Autor: Nathan A. Fischer, Malik Y. Kahook, Suhail Abdullah, Eric Porteous, David A. Ammar, Jennifer L. Patnaik, Jeffrey R. SooHoo
Jazyk: angličtina
Rok vydání: 2020
Předmět:
Zdroj: Ophthalmology and Therapy, Vol 9, Iss 2, Pp 279-291 (2020)
Druh dokumentu: article
ISSN: 2193-8245
2193-6528
DOI: 10.1007/s40123-020-00242-0
Popis: Abstract Purpose To quantify the effects of modified Ahmed glaucoma valves® (AGV) with anti-fibrotic plate coatings or a plate surface micro-pattern on outflow resistance and tissue response. Methods Twelve New Zealand rabbits were divided into four groups: commercially available AGV implants (n = 3), AGV with hydrophilic coating (n = 3), AGV with heparin coating (n = 3), and AGV with a plate surface micro-pattern (n = 3). After 6 weeks, the anterior chamber silicone tube was cannulated in situ and perfused with 2.5 μL/min of saline. The pressures were recorded with a perfusion system to measure outflow resistance. The rabbits were then euthanized followed by enucleation of all eyes for bleb histological analyses. Results Hydrostatic pressures were significantly lower in AGVs with the hydrophilic plate coating (mean difference −9.6 mm Hg; p
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje