Effect of Novel Design Modifications on Fibrotic Encapsulation: An In Vivo Glaucoma Drainage Device Study in a Rabbit Model
Autor: | Nathan A. Fischer, Malik Y. Kahook, Suhail Abdullah, Eric Porteous, David A. Ammar, Jennifer L. Patnaik, Jeffrey R. SooHoo |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2020 |
Předmět: | |
Zdroj: | Ophthalmology and Therapy, Vol 9, Iss 2, Pp 279-291 (2020) |
Druh dokumentu: | article |
ISSN: | 2193-8245 2193-6528 |
DOI: | 10.1007/s40123-020-00242-0 |
Popis: | Abstract Purpose To quantify the effects of modified Ahmed glaucoma valves® (AGV) with anti-fibrotic plate coatings or a plate surface micro-pattern on outflow resistance and tissue response. Methods Twelve New Zealand rabbits were divided into four groups: commercially available AGV implants (n = 3), AGV with hydrophilic coating (n = 3), AGV with heparin coating (n = 3), and AGV with a plate surface micro-pattern (n = 3). After 6 weeks, the anterior chamber silicone tube was cannulated in situ and perfused with 2.5 μL/min of saline. The pressures were recorded with a perfusion system to measure outflow resistance. The rabbits were then euthanized followed by enucleation of all eyes for bleb histological analyses. Results Hydrostatic pressures were significantly lower in AGVs with the hydrophilic plate coating (mean difference −9.6 mm Hg; p |
Databáze: | Directory of Open Access Journals |
Externí odkaz: | |
Nepřihlášeným uživatelům se plný text nezobrazuje | K zobrazení výsledku je třeba se přihlásit. |