Autor: |
Kewei Song, Kayo Hirose, Kioto Niitsu, Tsubasa Sui, Hiroto Kojima, Toshinori Fujie, Shinjiro Umezu |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Discover Nano, Vol 19, Iss 1, Pp 1-13 (2024) |
Druh dokumentu: |
article |
ISSN: |
2731-9229 |
DOI: |
10.1186/s11671-024-03988-9 |
Popis: |
Abstract Heart disease-related deaths have increased in recent decades, with most patients dying of sudden cardiac arrest. In such instances, the effect of regular electrocardiogram (ECG) measurements is minimal. Therefore, long-term ECG monitoring has become increasingly important. In this paper, we report a non-adhesive high accuracy ECG monitoring system that can be used in various scenarios without interfering with daily activities. The ECG ultra-thin film electrode is made by water-resistant material based on poly(3,4-ethylenedioxythiophene) poly(4-styrenesulfonate) (PEDOT: PSS) electrode doped with ethylene glycol (EG) and xylitol, to improve the noise signal caused by sweat. The optimal ratio of the three ingredients of PEDOT: PSS/xylitol/EG was determined experimentally to accommodate the ECG monitoring. By using the proposed selectively closed multi-channel single-lead logic circuit, the noise of ECG signal received from the proposed film electrode can be successfully reduced during broad-area electrode measurements, thus to improve ECG measurement accuracy. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|