Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements
Autor: | J. Saarelainen, M. Hakulinen, T. Rikkonen, H. Kröger, M. Tuppurainen, H. Koivumaa-Honkanen, R. Honkanen, M. Hujo, J. S. Jurvelin |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2016 |
Předmět: | |
Zdroj: | Journal of Osteoporosis, Vol 2016 (2016) |
Druh dokumentu: | article |
ISSN: | 2090-8059 2042-0064 |
DOI: | 10.1155/2016/1424582 |
Popis: | In long-term prospective studies, dual-energy X-ray absorptiometry (DXA) devices need to be inevitably changed. It is essential to assess whether systematic differences will exist between measurements with the new and old device. A group of female volunteers (21–72 years) underwent anteroposterior lumbar spine L2–L4 (n=72), proximal femur (n=72), and total body (n=62) measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) measurements with these two scanners showed a high linear association at all tested sites (r=0.962–0.995; p |