Two different autorefractors for vision screening in children and adolescents

Autor: Xiao-Jun Chen, Lin-Jie Liu, Bing Sun, Dan-Dan Jiang, Shu-Shu Zuo, Yan-Hui Wang, Chun-Hua Zhang, Yan-Yan Chen
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: International Journal of Ophthalmology, Vol 17, Iss 2, Pp 331-338 (2024)
Druh dokumentu: article
ISSN: 2222-3959
2227-4898
DOI: 10.18240/ijo.2024.02.16
Popis: AIM: To compare the consistency of two autorefractors (Tianle RM-9000 and Topcon KR-800) for school-age myopia children, and to provide a basis for largescale data analysis and comparison. METHODS: The refractive error in 909 subjects (age 4–18y) were measured using both autorefractors without cycloplegia. The data were analyzed using Fourier decomposition and the correlation coefficients, intraclass correlation coefficients (ICC), and Bland-Altman limits of agreement (LoA) for each parameter were calculated. RESULTS: There was a strong correlation between the spherical equivalent (SE), sphere diopter (DS), and cylinder diopter (DC) readings of the Tianle RM-9000 and those of the Topcon KR-800, with correlation coefficient values of 0.98, 0.98 and 0.83 and ICC values of 0.99, 0.99 and 0.93, respectively. However, the correlation coefficients and ICC values of J0 and J45 were unreliable (R=-0.004, -0.034; both ICC
Databáze: Directory of Open Access Journals