Late thrombosis of Gore Cardioform septal occluder device in a patient with history of patent foramen ovale closure

Autor: Muhammad Hasib Khalil, MBBS, Mobeen Haider, MBBS, Sanjay Mehta, MD
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Radiology Case Reports, Vol 19, Iss 4, Pp 1386-1390 (2024)
Druh dokumentu: article
ISSN: 1930-0433
DOI: 10.1016/j.radcr.2023.12.047
Popis: Patent foramen ovale (PFO) is a flaplike opening in the fossa ovalis; between the septum primum and secundum. It is highly prevalent with approximately 25% of the population having a PFO. It is usually asymptomatic but can rarely cause paradoxical embolism leading to stroke and/or significant right to left shunting causing hypoxia. The complications of PFO closure with a percutaneous device include embolization, cardiac perforation, and thrombosis. These are all early complications. We describe the case of a patient with a history of PFO closure who had device thrombosis 3 years after implantation. Management includes anticoagulation with warfarin. Repeat cardiac imaging to document the resolution of thrombosis is recommended.
Databáze: Directory of Open Access Journals