Insufficient procedural anticoagulation during leadless pacing led to catheter-related thrombosis in a hemodialysis patient

Autor: Qiang Chen, Yong Jiang Ma, Chun Hong Zhang, Li Wei Zhang
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: BMC Cardiovascular Disorders, Vol 21, Iss 1, Pp 1-5 (2021)
Druh dokumentu: article
ISSN: 1471-2261
DOI: 10.1186/s12872-021-02318-6
Popis: Abstract Background Leadless pacemaker was a promising innovation than traditional transvenous pacemaker, the procedural complications were prone to be bleeding-related. However, very few reports also concerned about the thrombus formation during the procedure. Case presentation A hemodialysis patient with diabetic gangrene of right foot suffered from catheter-related thrombosis during leadless pacing, resulting in failure of recapture the pacemaker. A low activated clotting time (ACT) level of 104 s confirmed the insufficiency of anticoagulation. Finally, the whole delivery catheter had to be removed from the delivery sheath, another new pacemaker system was applied and successfully implanted after adjusting the ACT level to 248 s. Conclusion Catheter-related thrombosis could be a large obstacle for leadless pacemaker implantation. In addition to routine anticoagulation, ACT monitoring might be necessary during the procedure.
Databáze: Directory of Open Access Journals