Popis: |
Background: Measuring the fraction of inspired oxygen (FiO2) is challenging in spontaneously breathing patients with impaired respiratory mechanics during low-flow nasal cannula. Our study investigates the FiO2 with varied tidal volume (VT) and respiratory rate (RR) among different lung mechanics and provides equations to estimate the FiO2. Methods: Two training and test lungs were used in this study, and the three lung mechanics (normal (R5/C60), restrictive (R20/C80), obstructive (R5/C40)) were designed. Spontaneous breathing with VT (300, 500, and 700 mL) and RR (10, 20, and 30 breaths/min) was simulated. The flow rate of the nasal cannula was set to 1, 3, and 5 L per minute (LPM), and the FiO2 was measured at the carina. Results: The lowest and highest FiO2 were evident during high (700 mL) and low VT (300 mL), respectively, among normal, restrictive, and obstructive lung models. As RR increases, this decreases the FiO2. However, we found that VT and oxygen flow rate are the principal factors influencing measured FiO2 by multiple linear regression analysis. Conclusions: Our data suggest that the actual FiO2 is never as high in spontaneously breathing patients as that estimated. VT and oxygen flow rate had a substantial impact on the FiO2. |