Autor: |
Yavuz Cakir, Antoine G. Sassine, Reem Amine, Karen Matar, Katherine E. Talcott, Sunil K. Srivastava, Jamie L. Reese, Justis P. Ehlers |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Scientific Reports, Vol 14, Iss 1, Pp 1-6 (2024) |
Druh dokumentu: |
article |
ISSN: |
2045-2322 |
DOI: |
10.1038/s41598-024-67700-8 |
Popis: |
Abstract This study aimed to assess retinal alterations following membrane peeling procedures using novel ILM Forceps with laser ablated surface with the help of intraoperative optical coherence tomography (iOCT). A post-hoc analysis was performed to evaluate iOCT findings in eyes that underwent membrane peeling procedures with the novel ILM Forceps. Pre-peel and post-peel iOCT videos and images were evaluated for each eye to assess for post-peel anatomic alterations. Surgical video/iOCT scan correlation was conducted to evaluate the etiology of anatomic alterations. Thirty-two eyes were included in the analysis. Three eyes (9%) had focal full thickness retinal elevations identified on iOCT following tissue-instrument interaction with the ILM Forceps. Two eyes (6%) had focal inner retinal elevations and one eye (3%) had a full-thickness retinal elevation that were not related to direct tissue-instrument interaction but rather indirect peeling forces. iOCT-identified architectural alterations related to direct-tissue instrument interaction were relatively infrequent ( |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |
K zobrazení výsledku je třeba se přihlásit.
|