Open-Path Cavity Ring-Down Spectroscopy for Simultaneous Detection of Hydrogen Chloride and Particles in Cleanroom Environment

Autor: Muhammad Bilal Khan, Christian L’Orange, Cheongha Lim, Deokhyeon Kwon, Azer P. Yalin
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Sensors, Vol 24, Iss 17, p 5611 (2024)
Druh dokumentu: article
ISSN: 1424-8220
DOI: 10.3390/s24175611
Popis: The present study addresses advanced monitoring techniques for particles and airborne molecular contaminants (AMCs) in cleanroom environments, which are crucial for ensuring the integrity of semiconductor manufacturing processes. We focus on quantifying particle levels and a representative AMC, hydrogen chloride (HCl), having known detrimental effects on equipment longevity, product yield, and human health. We have developed a compact laser sensor based on open-path cavity ring-down spectroscopy (CRDS) using a 1742 nm near-infrared diode laser source. The sensor enables the high-sensitivity detection of HCl through absorption by the 2-0 vibrational band with an Allan deviation of 0.15 parts per billion (ppb) over 15 min. For quantifying particle number concentrations, we examine various detection methods based on statistical analyses of Mie scattering-induced ring-down time fluctuations. We find that the ring-down distributions’ 3rd and 4th standard moments allow particle detection at densities as low as ~105 m−3 (diameter > 1 μm). These findings provide a basis for the future development of compact cleanroom monitoring instrumentation for wafer-level monitoring for both AMC and particles, including mobile platforms.
Databáze: Directory of Open Access Journals