Agreements’ profile of Scheimpflug-based optical biometer with gold standard partial coherence interferometry
Autor: | Sara Sardari, Akbar Fotouhi, Ebrahim Jafarzadehpur, Mehdi Khabazkhoob |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | International Journal of Ophthalmology, Vol 16, Iss 12, Pp 2095-2104 (2023) |
Druh dokumentu: | article |
ISSN: | 2222-3959 2227-4898 |
DOI: | 10.18240/ijo.2023.12.24 |
Popis: | AIM: To determine the agreement of ocular biometric indices including axial length, keratometric readings, anterior chamber depth, and horizontal corneal diameter between the Pentacam AXL and IOL Master 500. METHODS: The study was a large cross-sectional population-based study (Tehran Geriatric Eye Study) conducted from Jan 2019 to Jan 2020. A total of 160 clusters were randomly selected proportional to size (each cluster contained 20 individuals) from 22 strata of Tehran city. All people aged 60y and above were invited to participate in the study. For all participants, preliminary ocular examinations were performed including the measurement of uncorrected and best-corrected visual acuity, objective and subjective refraction, anterior and posterior segment examinations. All participants underwent an ocular biometry using the Pentacam AXL and IOL Master 500. RESULTS: The 95% limits of agreement (LoA) between the two devices were -0.13 to 0.19, -0.15 to 0.17, and -0.13 to 0.19 in normal, pseudophakic, and cataractous eyes, respectively. With increasing the axial length, the difference between the two devices significantly increased in all three groups of normal, pseudophakic, and cataractous eyes (P |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |