Impact of tricuspid annuloplasty device shape and size on valve mechanics—a computational studyCentral MessagePerspective

Autor: Collin E. Haese, MS, Mrudang Mathur, BTech, Chien-Yu Lin, MS, Marcin Malinowski, MD, PhD, Tomasz A. Timek, MD, PhD, Manuel K. Rausch, PhD
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: JTCVS Open, Vol 17, Iss , Pp 111-120 (2024)
Druh dokumentu: article
ISSN: 2666-2736
DOI: 10.1016/j.xjon.2023.11.002
Popis: Background: Tricuspid valve disease significantly affects 1.6 million Americans. The gold standard treatment for tricuspid disease is the implantation of annuloplasty devices. These ring-like devices come in various shapes and sizes. Choices for both shape and size are most often made by surgical intuition rather than scientific rationale. Methods: To understand the impact of shape and size on valve mechanics and to provide a rational basis for their selection, we used a subject-specific finite element model to conduct a virtual case study. That is, we implanted 4 different annuloplasty devices of 6 different sizes in our virtual patient. After each virtual surgery, we computed the coaptation area, leaflet end-systolic angles, leaflet stress, and chordal forces. Results: We found that contoured devices are better at normalizing end-systolic angles, whereas the one flat device, the Edwards Classic, maximized the coaptation area and minimized leaflet stress and chordal forces. We further found that reducing device size led to increased coaptation area but also negatively impacted end-systolic angles, stress, and chordal forces. Conclusions: Based on our analyses of the coaptation area, leaflet motion, leaflet stress, and chordal forces, we found that device shape and size have a significant impact on valve mechanics. Thereby, our study also demonstrates the value of simulation tools and device tests in “virtual patients.” Expanding our study to many more valves may, in the future, allow for universal recommendations.
Databáze: Directory of Open Access Journals