Impact of Aerobika® oscillating positive expiratory pressure in improving small airway resistance, lung function, symptoms and exercise capacity in chronic obstructive pulmonary disease

Autor: Siti Nurhanis Sahardin, Mas Fazlin Mohamad Jailaini, Nik Nuratiqah Nik Abeed, Andrea Yu-Lin Ban, Ng Boon Hau, Azat Azrai Azmel, Shamsul Azhar Shah, Mohamed Faisal Abdul Hamid
Jazyk: angličtina
Rok vydání: 2023
Předmět:
Zdroj: Frontiers in Medicine, Vol 10 (2023)
Druh dokumentu: article
ISSN: 2296-858X
DOI: 10.3389/fmed.2023.1202380
Popis: BackgroundAerobika® oscillating positive expiratory pressure (OPEP) device promotes airway clearance in many respiratory diseases. However, studies have yet to focus on its effectiveness in improving small airway resistance via impulse oscillometry (IOS) measurement in COPD subjects. We aim to evaluate the improvement of small airway resistance (via IOS), lung function (spirometry), exercise capacity [via 6-min walking test (6MWT)], symptoms [COPD assessment test (CAT)] and severe exacerbation events among COPD subjects using Aerobika® OPEP.MethodsThis was a prospective, single-arm interventional study among COPD subjects with small airway disease. Subjects were instructed to use twice daily Aerobika® OPEP (10 min each session); for 24 weeks; as an additional to standard therapy. IOS, spirometry, 6MWT, CAT score and severe exacerbation events were evaluated at baseline, 12 weeks and 24 weeks.ResultsFifty-three subjects completed the study. Aerobika® usage showed improvement of IOS parameters; e.g. measurement of airway resistance at 5 Hz (R5), cmH20/L/s, (12-week p = 0.008, 24-week p
Databáze: Directory of Open Access Journals