Covered SEMS failed to cure airway fistula closed by an amplatzer device

Autor: Huibin Lu, Yahua Li, Kewei Ren, Zongming Li, Juanfang Liu, Xuhua Duan, Jianzhuang Ren, Xinwei Han
Jazyk: angličtina
Rok vydání: 2023
Předmět:
Zdroj: BMC Pulmonary Medicine, Vol 23, Iss 1, Pp 1-7 (2023)
Druh dokumentu: article
ISSN: 1471-2466
DOI: 10.1186/s12890-023-02548-8
Popis: Abstract Background Airway fistula is a rare but threatening complication associated with high rates of morbidity and mortality. We report the experience of Amplatzer device application in airway fistulae that failed to be cured with a covered self-expandable metallic stent (SEMS). Materials and methods Patients who failed occlusion with a covered self-expandable metallic stent and received Amplatzer device placement from Jan 2015 to Jan 2020 were retrospectively enrolled. A total of 14 patients aged 42 to 66 years (55.14 ± 7.87) were enrolled in this study. The primary diseases, types of fistula, types of stents, duration, size of fistula, and follow-up were recorded. Results All 14 patients with airway fistula failed to be occluded with a covered metallic stent and received Amplatzer device placement. Among the 14 patients, 6 had BPF, 3 had TEF and 5 had GBF. The average stent time was 141.93 ± 65.83 days. The sizes of the fistulae ranged from 3 to 6 mm. After Amplatzer device placement, the KPS score improved from 62.14 ± 4.26 to 75.71 ± 5.13 (P