A Coil Constant Calibration Method Based on the Phase-Frequency Response of Alkali Atomic Magnetometer
Autor: | Han Yao, Danyue Ma, Junpeng Zhao, Jixi Lu, Ming Ding |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2019 |
Předmět: | |
Zdroj: | Photonic Sensors, Vol 9, Iss 2, Pp 189-196 (2019) |
Druh dokumentu: | article |
ISSN: | 1674-9251 2190-7439 |
DOI: | 10.1007/s13320-019-0530-4 |
Popis: | Abstract We propose an in-situ method to calibrate the coil constants of the optical atomic magnetometer. This method is based on measuring the Larmor precession of spin polarized alkali metal atoms and has been demonstrated on a K-Rb hybrid atomic magnetometer. Oscillation fields of different frequencies are swept on the transverse coil. By extracting the resonance frequency through phase-frequency analysis of electron spin projection, the coil constants are calibrated to be 323.1 ± 0.28 nT/mA, 108 ± 0.04 nT/Ma, and 185.8 ± 1.03 nT/mA along the X, Y, and Z directions, respectively. |