Popis: |
Background: Second generation extraglottic airway devices with gastric access and separate breathing channels have ushered in a new era where their use is increasingly prevalent in surgical patients who would have been traditionally intubated for general anesthesia. New innovations like the i-gel, which is constructed of a thermoplastic elastomer, provide an airtight seal around patient′s perilaryngeal anatomy without the inflatable cuff mechanism found in the laryngeal mask airway supreme (LMAS). Methods: We conducted a randomized controlled trial comparing the LMAS with the i-gel in 70 anesthetized paralyzed patients undergoing laparoscopic female sterilization. Our primary outcome measure was the oropharyngeal leak pressure (OLP). We studied secondary outcomes of successful first attempt insertion rates, time and ease of the airway and gastric tube insertion, leak fractions and pharyngeal morbidity. Results: We found no difference in the OLP between LMAS and i-gel, 25.9 (4.2) versus 24.4 (4.3) s, P=0.153. Both devices had similar first attempt insertion rates (LMAS 94% vs. i-gel 91%) with similar ease and comparable times to achieve an effective airway, LMAS 14.7 (2.7) versus i-gel 16.5 (9.6) s, P=0.306, although gastric tube insertion was easier and faster for the LMAS, 7.9 (1.9) versus i-gel 14.8 (7.7) s, P |