Volar Locking Plate Breakage after Nonunion of a Distal Radius Osteotomy
Autor: | Sergi Barrera-Ochoa, Sergi Rodríguez-Alabau, Andrea Sallent, Francisco Soldado, Xavier Mir |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2016 |
Předmět: | |
Zdroj: | Case Reports in Medicine, Vol 2016 (2016) |
Druh dokumentu: | article |
ISSN: | 1687-9627 1687-9635 |
DOI: | 10.1155/2016/6836190 |
Popis: | We report a 38-year-old male with a nonunion followed by plate breakage after volar plating of a distal radius osteotomy. Volar locking plates have added a new approach to the treatment of distal radius malunions, due to a lower morbidity of the surgical approach and the strength of the final construction, allowing early mobilization and return to function. Conclusion. Plate breakage is an uncommon complication of volar locking plate fixation. To our knowledge, few cases have been described after a distal radius fracture and no case has been described after a distal radius corrective osteotomy. In the present case, plate breakage appears to have occurred as a result of a combination of multiple factors as the large corrective lengthening osteotomy, the use of demineralized bone matrix instead of bone graft, and the inappropriate fixation technique as an unfilled screw on the osteotomy site, rather than the choice of plate. |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |