Visual field examination method using virtual reality glasses compared with the Humphrey perimeter

Autor: Tsapakis S, Papaconstantinou D, Diagourtas A, Droutsas K, Andreanos K, Moschos MM, Brouzas D
Jazyk: angličtina
Rok vydání: 2017
Předmět:
Zdroj: Clinical Ophthalmology, Vol Volume 11, Pp 1431-1443 (2017)
Druh dokumentu: article
ISSN: 1177-5483
Popis: Stylianos Tsapakis, Dimitrios Papaconstantinou, Andreas Diagourtas, Konstantinos Droutsas, Konstantinos Andreanos, Marilita M Moschos, Dimitrios Brouzas 1st Department of Ophthalmology, National and Kapodistrian University of Athens, Athens, Greece Purpose: To present a visual field examination method using virtual reality glasses and evaluate the reliability of the method by comparing the results with those of the Humphrey perimeter.Materials and methods: Virtual reality glasses, a smartphone with a 6 inch display, and software that implements a fast-threshold 3 dB step staircase algorithm for the central 24° of visual field (52 points) were used to test 20 eyes of 10 patients, who were tested in a random and consecutive order as they appeared in our glaucoma department. The results were compared with those obtained from the same patients using the Humphrey perimeter.Results: High correlation coefficient (r=0.808, P