Desflurane consumption during automated closed-circuit delivery is higher than when a conventional anesthesia machine is used with a simple vaporizer-O2-N2O fresh gas flow sequence

Autor: Sosnowski Maurice, Deloof Thierry, Carette Rik, Dewulf Bram BC, De Mey Nathalie, De Cooman Sofie, De Wolf Andre M, Hendrickx Jan FA
Jazyk: angličtina
Rok vydání: 2008
Předmět:
Zdroj: BMC Anesthesiology, Vol 8, Iss 1, p 4 (2008)
Druh dokumentu: article
ISSN: 1471-2253
DOI: 10.1186/1471-2253-8-4
Popis: Abstract Background The Zeus® (Dräger, Lübeck, Germany), an automated closed-circuit anesthesia machine, uses high fresh gas flows (FGF) to wash-in the circuit and the lungs, and intermittently flushes the system to remove unwanted N2. We hypothesized this could increase desflurane consumption to such an extent that agent consumption might become higher than with a conventional anesthesia machine (Anesthesia Delivery Unit [ADU®], GE, Helsinki, Finland) used with a previously derived desflurane-O2-N2O administration schedule that allows early FGF reduction. Methods Thirty-four ASA PS I or II patients undergoing plastic, urologic, or gynecologic surgery received desflurane in O2/N2O. In the ADU group (n = 24), an initial 3 min high FGF of O2 and N2O (2 and 4 L.min-1, respectively) was used, followed by 0.3 L.min-1 O2 + 0.4 L.min-1 N2O. The desflurane vaporizer setting (FD) was 6.5% for the first 15 min, and 5.5% during the next 25 min. In the Zeus group (n = 10), the Zeus® was used in automated closed circuit anesthesia mode with a selected end-expired (FA) desflurane target of 4.6%, and O2/N2O as the carrier gases with a target inspired O2% of 30%. Desflurane FA and consumption during the first 40 min were compared using repeated measures one-way ANOVA. Results Age and weight did not differ between the groups (P > 0.05), but patients in the Zeus group were taller (P = 0.04). In the Zeus group, the desflurane FA was lower during the first 3 min (P < 0.05), identical at 4 min (P > 0.05), and slightly higher after 4 min (P < 0.05). Desflurane consumption was higher in the Zeus group at all times, a difference that persisted after correcting for the small difference in FA between the two groups. Conclusion Agent consumption with an automated closed-circuit anesthesia machine is higher than with a conventional anesthesia machine when the latter is used with a specific vaporizer-FGF sequence. Agent consumption during automated delivery might be further reduced by optimizing the algorithm(s) that manages the initial FGF or by tolerating some N2 in the circuit to minimize the need for intermittent flushing.
Databáze: Directory of Open Access Journals