Ablation zone geometry after CT-guided hepatic microwave ablation: evaluation of a semi-automatic software and comparison of two different ablation systems
Autor: | Van Dai Vo Chieu, Frank Wacker, Christian Rieder, Gesa H. Pöhler, Christian Schumann, Hanne Ballhausen, Kristina I. Ringe |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2020 |
Předmět: | |
Zdroj: | International Journal of Hyperthermia, Vol 37, Iss 1, Pp 533-541 (2020) |
Druh dokumentu: | article |
ISSN: | 0265-6736 1464-5157 02656736 |
DOI: | 10.1080/02656736.2020.1766704 |
Popis: | Purpose The aims of this study were to evaluate a semi-automatic segmentation software for assessment of ablation zone geometry in computed tomography (CT)-guided microwave ablation (MWA) of liver tumors and to compare two different MWA systems. Material and Methods 27 patients with 40 hepatic tumors (primary liver tumor n = 20, metastases n = 20) referred for CT-guided MWA were included in this retrospective IRB-approved study. MWA was performed using two systems (system 1: 915 MHz; n = 20; system 2: 2.45 GHz; n = 20). Ablation zone segmentation and ellipticity index calculations were performed using SAFIR (Software Assistant for Interventional Radiology). To validate semi-automatic software calculations, results (2 perpendicular diameters, ellipticity index, volume) were compared with those of manual analysis (intraclass correlation, Pearson’s correlation, Mann–Whitney U test; p |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |