Investigation of Topographical Alterations in Titanium-Zirconium-Alloy Implant Threads following Er:YAG Irradiation

Autor: Mustafa Ayna, Johannes Spille, Yahya Açil, Jan-Tobias Weitkamp, Jörg Wiltfang, Cemal Esen, Aydin Gülses
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Materials, Vol 15, Iss 22, p 7889 (2022)
Druh dokumentu: article
ISSN: 1996-1944
DOI: 10.3390/ma15227889
Popis: The aim of the current experimental study was to comparatively assess the surface alterations in titanium and titanium-zirconium alloy implants in terms of thread pitch topography after irradiation with an Er:YAG laser, which is recommended in the literature for its sterilizing effect in the treatment of contaminated implant surfaces. Roxolid® and SLA® (Sand-blasted, Large-grit, Acid-etched) implants from Straumann® company with the same macro topography were investigated. The surface treatment was carried out using a wavelength of 2940 nm, 60 s irradiation time, a frequency of 10 Hz, and energies between 120 mJ and 250 mJ. The alterations were quantitatively analyzed by conducting roughness analysis via white light interferometry and qualitatively using SEM images. Roxolid® could particularly maintain its surface topography at a level of 160 mJ. At an energy level of 250 mJ, the surface properties of the pitch could be significantly altered for the first time. Compared to the Standard Plus dental implants studied, no distinct removal of the material from the surface was detected. The alloy properties of Roxolid® confirm the manufacturer’s statement in terms of stability and could offer advantages in peri-implantitis management if decontamination has been selected. However, as a part of a respective strategy, smoothening of a Roxolid® implant surface requires a significantly higher energy level compared to SLA-Standard® dental implants.
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje