Safety and efficacy of Amplatzer duct occluder II and konar-MF™ VSD occluder in the closure of perimembranous ventricular septal defects in children weighing less than 10 kg

Autor: Kaan Yildiz, Nazmi Narin, Sedef Oksuz, Rahmi Ozdemir, Ozge Pamukcu, Ali Baykan, Abdullah Ozyurt, Sedat Bagli, Rasit Aktas, Ikbal Nur Safak, Muhammed Akif Atlan, Yunus Sezer Bayam, Cem Karadeniz
Jazyk: angličtina
Rok vydání: 2023
Předmět:
Zdroj: Frontiers in Cardiovascular Medicine, Vol 10 (2023)
Druh dokumentu: article
ISSN: 2297-055X
DOI: 10.3389/fcvm.2023.1255808
Popis: IntroductionDevice closure of perimembranous ventricular septal defects (pmVSD) is a successful off-label treatment alternative. We aim to report and compare the outcomes of pmVSD closure in children weighing less than 10 kg using Amplatzer Duct Occluder II (ADOII) and Konar-MF VSD Occluder (MFO) devices.MethodsRetrospective clinical data review of 52 children with hemodynamically significant pmVSD, and sent for transcatheter closure using ADOII and MFO, between January 2018 and January 2023. Baseline, procedural, and follow-up data were compared according to the implanted deviceResultsADOII devices were implanted in 22 children with a median age of 11 months (IQR, 4.1–14.7) and weight of 7.4 kg (IQR, 2.7–9.7). MFO devices were implanted in 30 children with a median age of 11 months (IQR, 4.8–16.6) and weight of 8 kg (IQR, 4.1–9.6). ADOII were implanted (retrograde, 68.1%) in defects with a median left ventricular diameter of 4.6 mm (IQR, 3.8–5.7) and right ventricular diameter of 3.5 mm (IQR, 3.1–4.9) while MFO were implanted (antegrade, 63.3%) in defects with a median left ventricular diameter of 7 mm (IQR, 5.2–11.3) (p > 0.05) and right ventricular diameter of 5 mm (IQR, 2.0, 3.5–6.2) (p
Databáze: Directory of Open Access Journals