Autor: |
Satoshi Nakasone, MD, PhD, Mika Takaesu, MD, Masato Ishihara, MD, Masamichi Onaga, MD, Takahiro Igei, MD, Yoshihide Miyata, MD, Kotaro Nishida, MD, PhD |
Jazyk: |
angličtina |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Arthroplasty Today, Vol 16, Iss , Pp 46-52 (2022) |
Druh dokumentu: |
article |
ISSN: |
2352-3441 |
DOI: |
10.1016/j.artd.2022.04.004 |
Popis: |
Background: Accurate cup placement during total hip arthroplasty (THA) is difficult because the intraoperative pelvic position changes even in supine patient position. We developed a device known as HipPointer; it corrects pelvic rotation and creates a functional pelvic plane as a reference. The aim of this study was to determine the device placement accuracy and investigate causes of error. Material and methods: HipPointer was used for cup placement in 353 hips of 308 patients who underwent direct-anterior-approach THA in supine position. The mean age at surgery and body mass index were 63.9 (17-90) years and 24.9 (16.6-42.0) kg/m2, respectively. The mean observation period was 40.5 (12-73) months. To investigate the accuracy of HipPointer, preoperative planning and postoperative cup placement angles relative to the functional pelvic plane were evaluated using a three-dimensional analysis software, and absolute errors were determined. Results: The means ± standard deviations of radiographic inclination (RI) and radiographic anteversion (RA) were 40.2 ± 3.0° and 15.8 ± 3.6°, respectively. The absolute errors of RI and RA were 2.2 ± 2.0° and 2.7 ± 2.3°, respectively. The ratio of the cup placement angle for which both RI and RA are ≤10° in the target zone was 99% (350/353 hips), and the ratio of the absolute errors for which both RI and RA are ≤5° was 80.4% (284/353 hips). Conclusions: HipPointer is simple in structure, easy to use, and useful for direct-anterior-approach THA in supine position. It provides good cup placement accuracy. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|