Autor: |
Valerio Caputo, Cristina Bax, Luca Colantoni, Cristina Peconi, Andrea Termine, Carlo Fabrizio, Giulia Calvino, Laura Luzzi, Giorgia Gaia Panunzi, Claudia Fusco, Claudia Strafella, Raffaella Cascella, Luca Battistini, Carlo Caltagirone, Antonino Salvia, Giulia Sancesario, Emiliano Giardina |
Jazyk: |
angličtina |
Rok vydání: |
2021 |
Předmět: |
|
Zdroj: |
International Journal of Infectious Diseases, Vol 108, Iss , Pp 187-189 (2021) |
Druh dokumentu: |
article |
ISSN: |
1201-9712 |
DOI: |
10.1016/j.ijid.2021.04.048 |
Popis: |
Objectives: The present study compared the performance of the Lumipulse G Sars-CoV-2 Ag kit with the TaqPath COVID-19 RT-PCR CE IVD kit. Methods: The study was conducted on 4266 naso-oropharyngeal swabs. Samples were subjected to antigen RT-PCR tests for the detection of Sars-CoV-2 and related variants. Statistical analyses were conducted in R software. Results: We found 503 positives (including 138 H69-V70 deletion carriers) and 3763 negatives by RT-PCR, whereas 538 positives and 3728 negatives were obtained by antigen testing. We achieved empirical and binormal AU-ROCs of 0.920 and 0.990, accuracy of 0.960, sensitivity of 0.866, specificity of 0.973, positive and negative predictive values of 0.810 and 0.980. We obtained a positive correlation between viral loads and antigen levels (R2 = 0.81), finding a complete concordance for high viral loads (log10 copies/mL > 5.4). Antigen levels > 222 pg/mL were found to be reliable in assigning positive samples (p < 0.01). Concerning variant carriers, antigen test detected them with the same accuracy as other positive samples. Conclusions: Molecular and antigen tests should be evaluated regarding the prevalence of the area. In case of low prevalence, antigen testing can be employed as a first-line screening for the timely identification of affected individuals with high viral load, also if carriers of Sars-CoV-2 variants. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|