Graphite furnace atomic absorption spectrometry as a routine method for the quantification of beryllium in blood and serum

Autor: Brousseau Pauline, Fournier Michel, Stephan Chadi H, Sauvé Sébastien
Jazyk: angličtina
Rok vydání: 2008
Předmět:
Zdroj: Chemistry Central Journal, Vol 2, Iss 1, p 14 (2008)
Druh dokumentu: article
ISSN: 1752-153X
DOI: 10.1186/1752-153X-2-14
Popis: Abstract Background A routine method for the quantification of beryllium in biological fluids is essential for the development of a chelation therapy for Chronic Beryllium Disease (CBD). We describe a procedure for the direct determination of beryllium in undigested micro quantities of human blood and serum using graphite furnace atomic absorption spectrometry. Blood and serum samples are prepared respectively by a simple 8-fold and 5-fold dilution with a Nash Reagent. Three experimental setups are compared: using no modifier, using magnesium nitrate and using palladium/citric acid as chemical modifiers. Results In serum, both modifiers did not improve the method sensitivity, the optimal pyrolysis and atomization temperatures are 1000°C and 2900°C, respectively. In blood, 6 μg of magnesium nitrate was found to improve the method sensitivity. The optimal pyrolysis and atomization temperatures were 800°C and 2800°C respectively. Conclusion In serum, the method detection limit was 2 ng l-1, the characteristic mass was 0.22 (± 0.07) pg and the accuracy ranged from 95 to 100%. In blood, the detection limit was 7 ng l-1, the characteristic mass was 0.20 (± 0.02) pg and the accuracy ranged from 99 to 101%.
Databáze: Directory of Open Access Journals