Optimized reamer geometry for controlled reaming of the proximal femur

Autor: Markus Heinecke, Frank Layher, Georg Matziolis
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Scientific Reports, Vol 14, Iss 1, Pp 1-7 (2024)
Druh dokumentu: article
ISSN: 2045-2322
DOI: 10.1038/s41598-024-55067-9
Popis: Abstract Preparation of the femoral proximal medullary cavity by reaming is essential for intramedullary nail osteosynthesis and hip revision arthroplasty. The use of reamers sometimes exerts high torsional forces on the bone. Design and direction of rotation of the reamer are potential influencing factors. The aim of this biomechanical study is to evaluate the best combination of a right- or left-cutting reamer with a clockwise- or counterclockwise-rotating insert in terms of preparation and safety. Right- and left-cutting reamers with conical design were each introduced into five synthetic femurs in both clockwise and counterclockwise rotation with constant feed force. A specially constructed test system was used for this series of tests, with which the respective intramedullary channel were reamed step by step. This was then used to determine the required torque. In addition, the feed rate measurement was analyzed using a modified digital caliper. The feed rates of the reamers with rotation in the same direction as the cutting direction were significantly increased compared to rotation in the opposite cutting direction (CCRLC vs. CCRRC 76.8 ± 9.0 mm/s vs. 25.2 ± 8.3 mm/s and CRRC vs. CRLC 54.3 ± 12.3 mm/s vs. 19.3 ± 0.6 mm/s; p
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje