Glucose measurements in sheep using a long-term continuous glucose monitor designed for humans

Autor: Barbara Roqueto dos Reis, Ryan K. Wright, Riley Thompson, Nikki Tabatabai, Katherine Amirault, Sathya Sujani, Robin R. White
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Frontiers in Veterinary Science, Vol 11 (2024)
Druh dokumentu: article
ISSN: 2297-1769
DOI: 10.3389/fvets.2024.1458044
Popis: This study evaluated the feasibility of utilizing a continuous glucose monitors (CGM) designed for use in humans to measure glucose levels in sheep. Four Suffolk x Dorset sheep were fitted with jugular catheters and FreeStyle Libre 2 (Abbott®) glucose monitors. Glucose concentration from the CGM were compared with those from a glucometer and traditional assays during a hyperglycemic clamp, aiming to explore a broader range of physiological glucose concentrations in a controlled manner. Measurements were taken every five minutes during the infusion and every ten minutes post-infusion until baseline levels were restored. Relationships were analyzed using a linear mixed-effects model with glucometer readings as the response variables, GCM reading as fixed effects, and animal as random effect with significant level of p