Evaluating Mobius3D Dose Calculation Accuracy for Small-Field Flattening- Filter-Free Photon Beams
Autor: | Changhwan Kim PhD, Jihun Kim PhD, Young Kyu Lee PhD, Han-Back Shin PhD, Min Cheol Han PhD, Hojin Kim PhD, Jin Sung Kim PhD |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2022 |
Předmět: | |
Zdroj: | Technology in Cancer Research & Treatment, Vol 21 (2022) |
Druh dokumentu: | article |
ISSN: | 1533-0338 15330338 |
DOI: | 10.1177/15330338221141542 |
Popis: | Purpose: We aimed to investigate the dose calculation accuracy of Mobius3D for small-field flattening-filter-free x-rays, mainly utilized for stereotactic body radiation therapy (SBRT). The accuracy of beam modeling and multileaf collimator (MLC) modeling in Mobius3D, significantly affecting the dose calculation is investigated. Methods: The commissioning procedures of Mobius3D were performed for unflattened 6 MV and 10 MV x-ray beams of the linear accelerator, including beam model adjustment and dosimetric leaf gap (DLG) optimization. An experimental study with artificial plans was conducted to evaluate the accuracy of small-field modeling. The dose calculation accuracy of Mobius3D was also evaluated for retrospective SBRT plans with multiple targets. Results: Both studies evaluated the dose calculation accuracy through comparisons with the measured data. Relatively large differences were observed for off-axis distances over 5 cm and for small fields less than 1 cm field size. For the study with artificial plans, the maximum absolute error of 9.96% for unflattened 6 MV and 9.07% for unflattened 10 MV was observed when the field size was 1 cm. For the study with patient plans, the mean gamma passing rate with 3%/3 mm gamma criterion was 63.6% for unflattened 6 MV and 82.6% for unflattened 10 MV. The maximum of the average dose difference was −19.9% for unflattened 6MV and −10.1% for unflattened 10MV. Conclusions: The dose calculation accuracy uncertainties of Mobius3D for small-field flattening-filter-free photon beams were observed. The study results indicated that the beam and MLC modeling of Mobius3D must be improved for use in SBRT pretreatment QA in clinical practice. |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |