Autor: |
Nathan Nadia S, Verhey Janko F |
Jazyk: |
angličtina |
Rok vydání: |
2004 |
Předmět: |
|
Zdroj: |
BioMedical Engineering OnLine, Vol 3, Iss 1, p 32 (2004) |
Druh dokumentu: |
article |
ISSN: |
1475-925X |
DOI: |
10.1186/1475-925X-3-32 |
Popis: |
Abstract Background Finite element method (FEM) analysis for intraoperative modeling of the left ventricle (LV) is presently not possible. Since 3D structural data of the LV is now obtainable using standard transesophageal echocardiography (TEE) devices intraoperatively, the present study describes a method to transfer this data into a commercially available FEM analysis system: ABAQUS©. Methods In this prospective study TomTec LV Analysis TEE© Software was used for semi-automatic endocardial border detection, reconstruction, and volume-rendering of the clinical 3D echocardiographic data. A newly developed software program MVCP FemCoGen©, written in Delphi, reformats the TomTec file structures in five patients for use in ABAQUS and allows visualization of regional deformation of the LV. Results This study demonstrates that a fully automated importation of 3D TEE data into FEM modeling is feasible and can be efficiently accomplished in the operating room. Conclusion For complete intraoperative 3D LV finite element analysis, three input elements are necessary: 1. time-gaited, reality-based structural information, 2. continuous LV pressure and 3. instantaneous tissue elastance. The first of these elements is now available using the methods presented herein. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|