Autor: |
Koji Tsutsumi, MD, Tatsuru Niibori, MD, Keiichiro Katsumoto, MD |
Jazyk: |
angličtina |
Rok vydání: |
2005 |
Předmět: |
|
Zdroj: |
Journal of Arrhythmia, Vol 21, Iss 4, Pp 475-479 (2005) |
Druh dokumentu: |
article |
ISSN: |
1880-4276 |
DOI: |
10.1016/S1880-4276(05)80008-0 |
Popis: |
Given concerns regarding electrode fixation and risk of dislodgement, transvenous implantation of a tined endocardial electrode into the right atrium is considered difficult in patients who require permanent atrial pacing following cardiac surgery. Implantation of a tined endocardial electrode into the right atrium was performed intraoperatively for 3 patients who required implantation of a permanent atrial electrode during the cardiac operation. This technique yielded excellent results during the mean follow-up period of 65 months, with low stimulation thresholds (mean 2.85 P), sufficiently high sensing thresholds (mean 2.23 μJ) and stable lead impedances (mean 491.7 ohm). This technique offers a useful and secure method for patients with preoperative bradycardial arrhythmias who require implantation of a permanent atrial electrode during open-heart surgery. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|