Autor: |
Gerrit Frommeyer, Sven Zumhagen, Dirk G. Dechering, Robert Larbig, Markus Bettin, Andreas Löher, Julia Köbe, Florian Reinke, Lars Eckardt |
Jazyk: |
angličtina |
Rok vydání: |
2016 |
Předmět: |
|
Zdroj: |
Journal of the American Heart Association: Cardiovascular and Cerebrovascular Disease, Vol 5, Iss 3 (2016) |
Druh dokumentu: |
article |
ISSN: |
2047-9980 |
DOI: |
10.1161/JAHA.115.003181 |
Popis: |
BackgroundThe results of the recently published randomized SIMPLE trial question the role of routine intraoperative defibrillation testing. However, testing is still recommended during implantation of the entirely subcutaneous implantable cardioverter‐defibrillator (S‐ICD) system. To address the question of whether defibrillation testing in S‐ICD systems is still necessary, we analyzed the data of a large, standard‐of‐care prospective single‐center S‐ICD registry. Methods and ResultsIn the present study, 102 consecutive patients received an S‐ICD for primary (n=50) or secondary prevention (n=52). Defibrillation testing was performed in all except 4 patients. In 74 (75%; 95% CI 0.66–0.83) of 98 patients, ventricular fibrillation was effectively terminated by the first programmed internal shock. In 24 (25%; 95% CI 0.22–0.44) of 98 patients, the first internal shock was ineffective and further internal or external shock deliveries were required. In these patients, programming to reversed shock polarity (n=14) or repositioning of the sensing lead (n=1) or the pulse generator (n=5) led to successful defibrillation. In 4 patients, a safety margin of |