EOS® is reliable to evaluate spinopelvic parameters: a validation study

Autor: Mohammadreza Shakeri, Seyed Mani Mahdavi, Masih Rikhtehgar, Mohammad Soleimani, Hasan Ghandhari, Behnam Jafari, Seyedehsan Daneshmand
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: BMC Medical Imaging, Vol 24, Iss 1, Pp 1-9 (2024)
Druh dokumentu: article
ISSN: 1471-2342
DOI: 10.1186/s12880-023-01178-0
Popis: Abstract Background Sagittal and coronal standing radiographs have been the standard imaging for assessing spinal alignment. However, their disadvantages include distortion at the image edges and low interobserver reliability in some parameters. EOS® is a low-dose biplanar digital radiographic imaging system that can avoid distortion by obtaining high-definition images. Methods This study aimed to evaluate spinopelvic parameters in conventional lateral C1S1 upright radiographs and EOS® images and compare them. Patients with non-deformity changes were subjected to routine clinical examinations. Plain AP and lateral X-ray radiographs were obtained along the entire spine length. Patients were also referred for full-length EOS® of the spine. Thoracic Kyphosis (TK), Lumbar Lordosis (LL), Pelvic Tilt (PT), Sacral Slope (SS), Pelvic Incidence (PI), and Sagittal Vertical Axis (SVA) were measured in the two studies by an orthopedic surgeon and a radiologist using PACS software. Also, the orthopedic surgeon evaluated the studies again after two weeks. Intra- and inter-observer reliability was then assessed using the interclass correlation coefficient (ICC). Also, the coefficient of variation was used to assess intra- and inter-observer reliability. Bland-Altman plots were drawn for each parameter. Results The mean age was 48.2 ± 6.6 years. Among the 50 patients, 30 (60%) were female. The mean ICC for TK, LL, PT, SS, PI, and SVA in EOS® images are 0.95, 0.95, 0.92, 0.90, 0.94, and 0.98, respectively, and in C1S1 radiography images, it was 0.92, 0.87, 0.94, 0.88, 0.93, and 0.98, respectively which shows good to excellent results. The coefficient of variation for intraobserver reliability was relatively low (
Databáze: Directory of Open Access Journals