Autor: |
Freddy Somarriba, Helen Khoury, Vinícius Saito Monteiro De Barros, Norma Alejandra Roas Zúniga, Joselin Judith Meza Martínez, Rosnory de los Ángeles Domínguez Cano |
Jazyk: |
English<br />Portuguese |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Brazilian Journal of Radiation Sciences, Vol 10, Iss 2A (2022) |
Druh dokumentu: |
article |
ISSN: |
2319-0612 |
DOI: |
10.15392/bjrs.v10i2A.1817 |
Popis: |
The aim of this paper is to present the results obtained in the intercomparison exercise organized by the Ionizing Radiation Metrology Laboratory at the Federal University of Pernambuco (LMRI-DEN/UFPE) and the Radiation Physics and Metrology Laboratory (LAF-RAM) of the National Autonomous University of Nicaragua, Managua (UNAN-Managua). The comparison aimed to assess the technical capabilities of the Dosimetry Calibration Laboratory (LCD) belonging to LAF-RAM for the irradiation of personal dosimeters in terms of Hp(d) magnitudes with a 137Cs radiation beam. The BeO OSL dosimeters were shipped by LMRI-DEN/UFPE to LAF-RAM for irradiation in March 2020. The evaluation of the Z-Score value obtained in the intercomparison was satisfactory for all irradiations of the dosimeters in the Hp (d) magnitude. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|