Analog Pulse Shape Discrimination Based on Time Duration and Pulse Height

Autor: Harn R., Osovizky A., Kadmon Y., Manor A., Ghelman M.
Jazyk: angličtina
Rok vydání: 2021
Předmět:
Zdroj: EPJ Web of Conferences, Vol 253, p 11008 (2021)
Druh dokumentu: article
ISSN: 2100-014X
DOI: 10.1051/epjconf/202125311008
Popis: Pulse shape discrimination is a name of a group of techniques used to detect and distinguish between different types of radiation interactions. Analog pulse shape discrimination methods can be more suitable than digital methods, for high-speed scintillators both from rate and power consumption perspectives. Common analog discrimination methods are based on pulse-height and pulse-energy discrimination techniques. Other techniques rely on the time difference in the pulse width such as the ZeroCrossing methods. Neither of the above combine both amplitude and time methods. We present a novel analog pulse shape discrimination topology that combines both height and time domain. The topology is based on discrimination according to the pulse duration in time combined with compensation function of the pulse height. Amplitude of the pulse is used as a restraining factor. Subsequently, our topology correctly identifies fast pulses that are prolonged in time due to their high amplitude. The topology yields superior discrimination capabilities, under degraded light collection conditions, with an uncertainty gap smaller than 1 ns in pulse width. The ability to control both the time and the amplitude parameters individually, provides tailored adjustment for various detectors and pulse shape discrimination applications.
Databáze: Directory of Open Access Journals