Autor: |
David M. Freccero, MD, Kenneth J. McAlpine, MD, Eric L. Smith, MD |
Jazyk: |
angličtina |
Rok vydání: |
2020 |
Předmět: |
|
Zdroj: |
Arthroplasty Today, Vol 6, Iss 2, Pp 210-213 (2020) |
Druh dokumentu: |
article |
ISSN: |
2352-3441 |
DOI: |
10.1016/j.artd.2020.01.011 |
Popis: |
We report a case of head-neck taper fretting corrosion in a patient who had a total hip replacement with a noncemented Stryker Anato femoral stem and a V40 metal head with a Stryker Tritanium hemispherical socket with a highly cross-linked polyethylene liner (metal on polyethylene) (Stryker, Mahwah, NJ, USA). A 57-year-old man presented with early-onset hip pain after right total hip arthroplasty. Workup was negative for infection. Metal artifact reduction sequence MRI revealed an encapsulated fluid mass. Metal ion cobalt level was elevated at 6 ppb. The patient underwent right revision total hip arthroplasty with excellent results at 1-year follow-up. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|