Autor: |
Mitsunari Maruyama, MD, PhD, Hisatoshi Araki, MD, Rika Yoshida, MD, PhD, Shinji Ando, MD, Megumi Nakamura, MD, PhD, Takeshi Yoshizako, MD, PhD, Yasushi Kaji, MD, PhD |
Jazyk: |
angličtina |
Rok vydání: |
2023 |
Předmět: |
|
Zdroj: |
Radiology Case Reports, Vol 18, Iss 12, Pp 4231-4234 (2023) |
Druh dokumentu: |
article |
ISSN: |
1930-0433 |
DOI: |
10.1016/j.radcr.2023.08.109 |
Popis: |
A high-flow steerable microcatheter has been reported to be useful as a triaxial system. Moreover, the benefits of steerable microcatheters in acute-angle bifurcation vessel insertions and a compact coil-packing technique using intentional folding with a bendable catheter tip have been reported. However, research on the usefulness of a high-flow steerable catheter and 0.025-inch guidewire combination (steerable-wire) technique is lacking. Herein, we report a case of balloon-occluded retrograde transvenous obliteration (BRTO) via the femoral venous approach to illustrate the usefulness of the steerable-wire technique. The steerable-wire technique facilitates the selection of the wire into the target vessel. The steerable-wire can be used instead of the 0.035-inch guidewire, which is versatile as other devices can follow the steerable wire. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|