Autor: |
Shah Urmi |
Jazyk: |
angličtina |
Rok vydání: |
2006 |
Předmět: |
|
Zdroj: |
Indian Journal of Ophthalmology, Vol 54, Iss 4, Pp 277-278 (2006) |
Druh dokumentu: |
article |
ISSN: |
0301-4738 |
Popis: |
The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure. The chip was successfully removed intraoperatively with a McPherson′s forceps. However, six months postoperatively patient complained of a disturbing glare due to posterior capsular folds originating from the site of IOL defect. Patient was advised, however refused to undergo Nd:Yag laser capsulotomy to relieve the symptoms. This case highlights the necessity to check each foldable IOL under the microscope before loading it into the cartridge, for possible edge defects. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|