Comparisons of an automated oscillometric device with a hybrid manual auscultatory device for the Korea National Health and Nutrition Examination Survey
Autor: | Yu‐Mi Kim, Sang Hyeon Park, Jinho Shin, Ki‐Chul Sung, Sang‐Hyun Ihm, Seong Heon Kim, Dae‐Hee Kim, Sang Min Park, In Jeong Cho, Kyung Won Oh, Eun Mi Lee |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2024 |
Předmět: | |
Zdroj: | The Journal of Clinical Hypertension, Vol 26, Iss 5, Pp 532-542 (2024) |
Druh dokumentu: | article |
ISSN: | 1751-7176 1524-6175 |
DOI: | 10.1111/jch.14805 |
Popis: | Abstract This study evaluated an oscillometric device (OD), Microlife WatchBP Office AFIB, and a hybrid manual auscultatory device (AD), Greenlight 300TM, to determine a suitable blood pressure (BP) measurement device for the Korea National Health and Nutrition Examination Survey in a mercury‐free context. Adhering to the 2018 Universal Standard's suggested consensus, the study involved 800 subjects (mean age 51.2 ± 17.5 years; 44.3% male), who underwent triplicate BP measurements following 5 min of rest in a randomized order (OD‐first: 398 participants; AD‐first: 402 participants). BP difference was calculated as OD value minus AD value, with results stratified by measurement sequence. The overall BP difference and tolerable error probability were ‐1.1 ± 6.5/‐2.6 ± 4.9 mmHg and 89.2%/92.5% for systolic/diastolic BP (SBP/DBP), respectively. Lin's concordance correlation coefficient was 0.907/0.844 for SBP/DBP (OD‐first/AD‐first: 0.925/0.892 for SBP, 0.842/0.845 for DBP). The overall agreement for hypertension (BP ≥ 140 and/or 90 mmHg) was 0.71 (p |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |