Autor: |
Chung-Ho Lee, Huiluo Cao, Shuo Jiang, Tammy Ting-Yan Wong, Cindy Wing-Sze Tse, Pak-Leung Ho |
Jazyk: |
angličtina |
Rok vydání: |
2024 |
Předmět: |
|
Zdroj: |
Diagnostics, Vol 14, Iss 12, p 1274 (2024) |
Druh dokumentu: |
article |
ISSN: |
2075-4418 |
DOI: |
10.3390/diagnostics14121274 |
Popis: |
The NG-Test CARBA 5 and Carbapenem-resistant K.N.I.V.O. Detection K-Set are lateral flow assays (LFAs) that rapidly detect five carbapenemases (KPC, NDM, IMP, VIM and OXA-48-like). We evaluated the effect of inoculum size on the performance of these two assays using 27 Enterobacterales isolates. Whole-genome sequencing (WGS) was used as the reference method. Using the NG-Test CARBA 5, eight Serratia spp. and six M. morganii isolates showed false-positive NDM results with a high inoculum. Using the Carbapenem-resistant K.N.I.V.O. Detection K-Set, eight M. morganii, four Serratia spp. and one K. pneumoniae isolates showed false-positive NDM and/or OXA-48-like bands at large inoculum sizes, while the other two M. morganii isolates demonstrated false-positive NDM and OXA-48-like results at all inoculum sizes. The false-positive bands varied in intensity. WGS confirmed that no carbapenemase gene was present. No protein sequence with a ≥50% identity to NDM or OXA-48-like enzymes was found. This study emphasizes the importance of assessing inoculum size in the diagnostic evaluation of LFAs. |
Databáze: |
Directory of Open Access Journals |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |