Adsorption of desflurane by the silica gel filters in breathing circuits: an in vitro study
Autor: | Seok Young Song, Bo Reum Lim, Taeha Ryu |
---|---|
Jazyk: | angličtina |
Rok vydání: | 2015 |
Předmět: | |
Zdroj: | Korean Journal of Anesthesiology, Vol 68, Iss 3, Pp 274-280 (2015) |
Druh dokumentu: | article |
ISSN: | 2005-6419 2005-7563 |
DOI: | 10.4097/kjae.2015.68.3.274 |
Popis: | BackgroundDuring general anesthesia, a heated breathing circuit (HBC) is used to replace the heat and moisture exchange function of the upper airway. One HBC uses an air dryer filter that employs silica gel (SG) as a desiccant. SG is capable of adsorbing many organic compounds. Therefore, we undertook an in vitro study of the adsorption of desflurane by SG filters.MethodsAn HBC was connected to an anesthesia machine, and a test lung was connected to the circuit. The test lung was mechanically ventilated with 2 or 4 L/min of fresh gas flow, with and without the air dryer filter. Desflurane was administered at a 6 vol% on the vaporizer dial setting. The experiment was repeated 15 times in each group. The end-tidal concentrations were measured during the experiments. The air dryer filter weights were measured before and after the experiments, and the times required to achieve the specific end-tidal desflurane concentrations were determined.ResultsSignificant differences in the end-tidal concentrations of desflurane were observed between the control and filter groups (P < 0.001). The filter weights increased significantly after the experiments (P < 0.001). The times required to achieve the same end-tidal desflurane concentrations were different with the application of the air dryer filter (P < 0.001).ConclusionsThe adsorption of desflurane with the use of an air dryer filter was verified in this in vitro study. Careful attention is needed when using air dryer gel filters during general anesthesia. |
Databáze: | Directory of Open Access Journals |
Externí odkaz: |