Visual field assessment in glaucoma: Comparative evaluation of manual Kinetic Goldmann Perimetry and Automated Static Perimetry

Autor: Agarwal Harish, Gulati V., Sihota Ramanjit
Jazyk: angličtina
Rok vydání: 2000
Předmět:
Zdroj: Indian Journal of Ophthalmology, Vol 48, Iss 4, Pp 301-6 (2000)
Druh dokumentu: article
ISSN: 0301-4738
27489884
Popis: Purpose: To compare the detection and assessment of progression of visual field defects in primary open-angle glaucoma with manual suprathreshold perimetry on Goldmann perimeter and automated static threshold perimetry on Humphery visual field (HVF) analyzer. Methods: 105 eyes of 54 patients of primary open-angle glaucoma were followed up with 3-monthly perimetry on Goldmann perimeter and HVF analyzer, for a period of 9 months. Results: HVF analyzer picked up visual field defects in 48 (46%) eyes whereas Goldmann perimeter picked up visual field defects in 26 (25%) eyes. HVF analyzer demonstrated progression in 14 eyes whereas Goldmann perimeter detected progression in 7 eyes during follow up of 9 months. Conclusions: HVF analyzer is superior to Goldmann perimeter to document and to demonstrate progression of visual field defects in primary open-angle glaucoma.