Assessment of Socket Pressure during Walking in Rapid Fit Prosthetic Sockets

Autor: Kazuhiko Sasaki, Gary Guerra, Win Lei Phyu, Sirarat Chaisumritchoke, Prawina Sutdet, Sirintip Kaewtip
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Sensors, Vol 22, Iss 14, p 5224 (2022)
Druh dokumentu: article
ISSN: 1424-8220
DOI: 10.3390/s22145224
Popis: (1) Background: A sustainable casting system that combines the use of a polystyrene bag, a prosthetic liner and a vacuum system was developed to reduce fabrication time while maintaining comfort for the trans-tibial prosthesis user. (2) Methods: Eight prosthetists (28.7 ± 8.25 years old) fit ten trans-tibial prosthesis wearers (46 ± 12.4 years old) with two types of total surface bearing (TSB) prostheses; a polystyrene bead (PS) prosthesis and a plaster of paris (POP) prosthesis. Duration of casting and combined mean peak pressure was measured at six locations on the residual limb using Force Sensing Resistors (FSR). A pressure uniformity score (%) was determined. Socket Comfort Scale (SCS) was also measured. (3) Results: Duration of casting for the POP method was 64.8 ± 9.53 min and 7.8 ± 2 min for the PS method, (p = 0.006). Pressure uniformity in the POP prosthesis was 79.3 ± 6.54 and 81.7 ± 5.83 in the PS prosthesis (p = 0.027). SCS in both prosthesis types were equivalent. (4) Conclusion: A rapid fit PS prosthesis was developed, with significantly shorter duration than the traditional POP method. Socket pressure uniformity was confirmed and improved in the PS method. Socket comfort was equal between the two prothesis types.