Autor: |
Anum Javed, Muhammad Owais Arshad, Javeria Nasir, Muhammad Hanif Chatni, Bushra Ayub |
Jazyk: |
angličtina |
Rok vydání: |
2022 |
Předmět: |
|
Zdroj: |
Journal of the Pakistan Medical Association, Vol 72, Iss 01 (2022) |
Druh dokumentu: |
article |
ISSN: |
0030-9982 |
DOI: |
10.47391/JPMA.11-1978 |
Popis: |
Objective: Glaucoma is a sight threatening disorder. Measuring the IOP with accuracy plays a pivotal role in diagnosing and monitoring glaucoma. The purpose of this study was to compare the values of non-contact tonometer; air-puff and GAT (Goldmann Applanation Tonometer) along with its correlation to central corneal thickness (CCT), in various IOP (intra-ocular pressure) groups. Methods: Cross-sectional, prospective study. 311 patients were enrolled in this study. IOP was taken in all patients with APT and GAT. CCT was also measured. Data was analyzed and correlated with the help of Bland-Altman, Pearson correlation and intraclass correlation analysis regarding APT, GAT and CCT, using SPSS 24.0 software. Results: The median IOP measured by APT and GAT was 14mmHg (range: 37) and 12mmHg (range: 16) where as the median CCT was 534µ (range: 44), respectively. At low ( |