Trueness of intraoral scanners according to subgingival depth of abutment for fixed prosthesis

Autor: Young-Tak Son, KeunBaDa Son, Kyu-Bok Lee
Jazyk: angličtina
Rok vydání: 2022
Předmět:
Zdroj: Scientific Reports, Vol 12, Iss 1, Pp 1-10 (2022)
Druh dokumentu: article
ISSN: 2045-2322
DOI: 10.1038/s41598-022-23498-x
Popis: Abstract This study aimed to compare the trueness of intraoral scanners (IOSs) according to the subgingival finish line depth of tooth preparation for fixed prostheses. The prepared maxillary right first molar was fabricated by using ceramic material. A computer-aided design (CAD) reference model (CRM) of the abutment was obtained by using a contact scanner. The subgingival finish line was located according to the depth at 0-mm, 0.25-mm, 0.5-mm, 0.75-mm, and 1-mm. CAD test models (CTMs) were obtained by using 2 IOSs (i500 and CS3600). CRM and CTM were superimposed and analyzed (Geomagic control X). The one-way analysis of variance (ANOVA) was used to compare the trueness according to the subgingival finish line depth. The paired t test was used to compare the trueness of IOSs with and without gingival retraction (α = .05). When the gingival displacement code was not used, it was observed that the trueness of both IOSs decreased significantly as the depth of the subgingival finish line increased (P
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje