Fast, Simple, and Sensitive Voltammetric Measurements of Acyclovir in Real Samples via Boron-Doped Diamond Electrode

Autor: Damian Gorylewski, Katarzyna Tyszczuk-Rotko, Magdalena Wójciak, Ireneusz Sowa
Jazyk: angličtina
Rok vydání: 2024
Předmět:
Zdroj: Materials, Vol 17, Iss 18, p 4480 (2024)
Druh dokumentu: article
ISSN: 1996-1944
DOI: 10.3390/ma17184480
Popis: The voltammetric acyclovir (ACV) trace-level determination procedure has been introduced. This is the first time that a commercially available boron-doped diamond electrode (BDDE) coupled with differential-pulse voltammetry (DPV) has been used for this purpose. The commercially available BDDE is characterized by a short response time, low background current, and very good analytical parameters of ACV determination. Ultimately, DPV measurements using the BDDE in 0.075 mol L−1 PBS with a pH of 7.2 under optimized conditions achieved the lowest detection limit (LOD = 0.0299 nmol L−1) reported in the literature for voltammetric procedures. Moreover, it is highly resistant to the presence of various interfering agents and has been used to analyze pharmaceutical and municipal wastewater samples. The obtained results are consistent with measurements made using chromatographic reference methods.
Databáze: Directory of Open Access Journals
Nepřihlášeným uživatelům se plný text nezobrazuje